Details
Tablet Hardness Tester - (TH 12 SMART)
Tablet Hardness Tester TH12 Smart is equipped with 7” High-resolution Display with Capacitive Touch Screen along with User interactive software for ease of operation with protection against invalid entries. The Tablet Hardness Tester – (TH 12 SMART) can perform an automatic run of up to 1-100 tablets, in addition to this, the run can be resumed even if it is interrupted by any power failure. Pause & Resume feature gives the user time to adjust the the tablets.TH 12 Smart tablet hardness tester has a multilevel passing limit to detect the values of Hardness, Length / Diameter, Thickness and Width with marking T1, T2, T3 for displaying the out of limit values. A complete 21 CFR part 11 complaint system with facilities like Audit trail, E Signature, multilevel password etc is inbuilt. For the purpose of safety, user authentication is performed for every operation done by the user of this hardness tester.
Standard Features of Tablet Hardness Tester - (TH 12 SMART)
- 7” High resolution Display with Capacitive Touch Screen & User interactive software for ease of operation with protection against invalid entries.
- Facility for date and time display with Built-in Real Time Clock (RTC)
- 12 Position Carousal for Automatic measurement of tablet parameters
- Thickness, Length / Diameter, Weight, Width &. Hardness
- Automatic run can be performed continuously up to 1-100 tablets. Run can be restarted incase of of interruption due to power failure
- Alphanumeric entries of Sample Name and Identification/Batch Number for authentication
- Daily Auto Incremented Run Number for authentication of each test Run
- Factory entered CUSTOMER NAME, Equipment ID with Instrument Serial Number on report
Process Stage,Press Details, comment entry with Identification / Batch No for additional information
Multilevel Passing Limit for detected values of Hardness, Length / Diameter, Thickness and Width with markingT1,T2,T3 for out of limit values - Calibration for Hardness, Thickness & Length / Diameter/Width :
- Hardness Calibration & Verification - 3 Point.
- Thickness Calibration & Verification - 3 Point.
- Length/Diameter/Width Calibration with 2 point & verification with 3 point.
- Calibration reminder function
- Quick RUN for Width, Thickness, Length/Diameter and Hardness measurement or All.
- The built in Sweeping blades cleans the tablet path for removal of broken pieces, debris & dust.
- Tablet Auto centering for length measurement as per tablet width / diameter.
- Groove for better positioning of oblong shape tablets.
- ‘Run’ halt feature for out of limit value with buzzer alarm & display message indication.
- ‘Pause’ / ‘Resume’ feature added while measuring Hardness to make a proper alignment of the tablet.
- Error indication helps user to trace the problem.
- Linear Displacement Method with selectable speed 0.2 to 1.6 mm/sec (Step of 0.2 mm/sec)
- Fast Advance and Return movement of breaking arm saves test run time.
- Result resolution 0.1N, 0.01mm and as per unit selected
- Audit Trail for all activities with search facility, report generation and printing
- Practically unlimited User ID’s with alphanumeric entries of user name, password and role based privileges selection
- Multi-level roles with password protection, expiry and complexity
- User authentication is performed for each and every operation done by user
- Reports can be saved in PDF format
- USB/Wifi/LAN Printing eliminates the need of serial port / parallel port to connect with instrument. The user can take printout on any android compatible local or network printer as well.
- Electronic signature functionality
- Data Backup & Restore facility available.
- Automatic backup of every run report (PDF) immediately after PDF generation (Optional)
- LAN (Ethernet) connectivity for Data backup on Network shared folder
Typical Specification
Parameters | Specifications |
Control | Micro controller based (Advanced version of microprocessor). |
Display | 7" High Resolution Display with Capacitive touch screen |
Test Parameter: |
No. of Tablets Practically Unlimited Tablets can be checked in a single run. |
1. Thickness: |
• Measuring Sensor - Linear Transducer • Measuring Range - 0.1mm to 15.0mm • Measuring Unit - mm, cm or inch • Measuring Accuracy - better than ± 0.06mm |
2. Width (in Separate Mode) : |
• Measuring Sensor - Stepper controlled Load Touch sensor • Measuring Range - 0.1mm to 15.0mm • Measuring Unit - mm, cm, inch • Measuring Accuracy - better than ± 0.06mm |
3. Diameter/ Length |
• Measuring Sensor - Stepper controlled Load Touch sensor • Measuring Range - Diameter 0.1mm to 25.0mm Length 0.1mm to 32.0mm(check the format) • Measuring Unit - mm, cm, inch • Measuring Accuracy - better than ± 0.06mm |
4. Hardness: Speed Control by |
Stepper motor • Linear Displacement 0.2, 0.4, 0.6, 0.8, 1.0, 1.2, 1.4, 1.6 mini sec with fast advance and return of breaking arm. • Measuring Sensor- Load Cell - Strain gauge • Measuring Range- 2N to 500N {SOON Optional) • Measuring Unit - N (Newton's) I Kp (Kilo Ponds) I Kgf{Sc - Strong Cobb) I Lbs (Pound) I UD (User Defined) • Measuring Accuracy ± 1N |
Weight: |
Manual Entry by User (for reporting purpose) Automatic Balance Interface - Sartorious, Mettler,etc when Automatic Wt. Transfer option is selected |
Method Storage: | Practically Unlimited Methods with Parameter |
Data Storage: | External Memory Card {16GB) (Expandable upto 32GB) |
Report Format: |
*GLP & Pharmacopoeia compliant report with Statistics (Avg, Min, Max and RSD) *Calibration Report *Audit Trial Reports |
Output: |
RS232 C ports for-Balance connectivity USB Port for--Print, Backup, Restore, Software Upgrade LAN Port for-- Print, Backup, Restore |
Instrument weight: | 16kgs Approx. |
Dimensions: | 340W x 270H x 4100 mm |
Power: | 110/220Vac, + 10%, 60 / 50Hz, 50 Watts, Option Available |
Environmental Operating Conditions: |
a) Operation : Indoor b) Temperature : Ambient to 45c c) Humidity : 5 to 90% non – condensing d) Altitude : upto 2000m |