Tablet Disintegration Tester DT 1000+

The Tablet Disintegration tester DT 1000+ is the advanced version where the baskets are driven individually and hence two different sets of samples can be run simultaneously. The time for individual tablets can be recorded. External temperature sensors check and display the temperature for each vessels. In Disintegration tester DT 1000+ there is also a provision to identify the position of disintegrated tablets with laser marked numbers on basket assembly.

At the end of the run the baskets automatically park out of the media. 20 methods can be stored with two tiers (Admin and User). Data can be downloaded to a PC or printed via serial port or parallel port.

Standard Features of Tablet Disintegration Tester 1000+
  • Meets current specification of USP, IP, BP and other Pharmacopoeias.
  • In DT1000+, Dual timer (baskets will have individual timer) and dual drive (each basket is driven by individual drive), 20*4 line display with pause and resume functionality to monitor disintegration in between the test.
  • Specifically designed for use in the quality and production control of normal plain coated tablets & gelatin capsules.
  • Programmable Temperature & Time.
  • Molded clear acrylic water bath with illumination for better visibility.
  • Auto Park out facility for basket after run completion.
  • Capability for dual buffer disintegration test.
  • Audible, low water level alarm with indication on display for safety.
  • External temperature sensors available for individual beaker.
  • Provision to identify the position of disintegrated tablet with laser marked numbers on basket assembly.
  • Provision to lock individual tablet disintegration time by using external key board with basket numbers.
  • PC based Data downloading facility using EDC 21 Software over LAN (Optional)

Typical Specification
 
Parameters
Specifications
 Drive
Dual Drive for 2 baskets with individual timer
Dip Speed
30, ±1 DPM
Stroke length
5.5, ± 0.1 cm
Water bath temperature range
Ambient +5 to 50°C
Water bath temperature accuracy
± 0.2°C
Water bath circulation
Immersion Pump
Beaker volume
1000 ml
Basket
 2X6 test positions (10 mesh) USP type A
Fluted Disk
6 numbers in each basket USP type A
Test Mode
Programmable Time up to 99:59 (hr:min)
Display Format 99:59:59
Display
20 X 4 Line back lighted LCD display
Keyboard
Alphanumeric splash water proof polyester soft keys
Output                                 Printer
                                            RS-232C
Parallel Port
PC Connectivity for data downloading
Power Requirement
230VAC / 50 Hz, 275 W
Environmental Operating conditions  
 a)  Operation Indoor
 b)  Temperature Ambient to 45°C
 c)  Humidity 20% to 80%
Dimension (Approx.)
330 X 330 X 580 mm (W X D X H)
Shipping Weight (Approx.)
16 kg.

Optional Accessories
  • Bolus Basket with 3 tubes & 1 tubes

  • Fluted Discs (For Above)