Genie U Water System
Specifications
Details
Genie U Ultrapure & RO Lab Water System produces ultrapure water (Type I water) and RO water from tap water directly. It offers desired solutions for research professionals who work with varieties of applications utilizing Type I ultrapure and RO pure water in the lab. Quality of ultrapure water meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards.
The system is CE/RoHS certified. RephiLe provides full documents support to meet user’s GMP, GSP, GAP, GCP, GLP requirements.
Features:
Design & Functions
- The system can be linked to multiple dispensers via CAN cable or wireless. “1+N” Mode.
- Automatic temperature compensation allows RO production rate stable over a wide range of temperature.
- Fully automated control system has cleaning, rising, flush and many other functions.
- System automatically rinses a new RO membrane. Users do not need to set special rinse program for a new RO membrane.
- A bypath automatically sends the RO water to the drain if it does not meet a preset quality standard. The RO drain water is reused to increase the water yield thus it is more environmentally friendly.
- The main system can be tucked away to save precious bench space since it fits well to be wall-mounted, set on the shelf or under the sink.
- The Chlorine cleaning and pH cleaning functions keep the maintenance easier and ensure optimal water quality.
- Leak protector incorporated can prevent damage from a water leakage by shutting the feed water off in case of unexpected leaking.
- Tank recirculation model guarantees water quality in the tank.
Effective TOC level controlment
- Built-in 185/254 nm dual wavelength lamp reduces TOC level in ultrapure water.
- A tank Sanitization Module is equipped to effectively keep microbial growth at a lower level by LED UVC light.
- On-line TOC monitor based on complete oxidation methods (for TOC models).
Easy Operation and Maintenance
- Performance history, alarms and alerts information, maintenance data of consumables as well as key parts are logged and traceable from a simple RFID scan at installation.
- RFID tags ensure perfect placement of consumables and trace their performance.
- Water quality, operation parameters, the status of the system, dispensers, components, and peripheral devices are stored and displayed on the large color touch screen monitor.
- Signature verification for maintenance and service.
Cartridges and Parts
1. Cartridge
- Pre-filtration pack removes most large particles, chlorine, calcium or organic compounds to protect RO membrane to extend their usable life..
- AC Pack contains special activated carbon to remove organic impurities and free chlorines.
- Ultra purification cartridge is filled with LeFilTM and OrganeFilTM media to remove trace ions and organics.
2. Control console
- 8-inch console controls system and peripheral devices (Such as leak protector and tank sanitization module). All operations can be done on the console by touching the screen.
- The control console can be placed on the bench or inside a drawer for further flexibility and space saving.
- The console screen and dispenser screens are water-drop proof. You can operate the console and dispenser with latex gloves on.
- The system has two level password protection on display menu for safety purposes.
3. Dispenser
- The volumetric function can automatically deliver the water volume you need once it is set up.
- With the adjusting button on a control console or a dispenser, you can dispense water at the flow rate you need up to 2 liters per minute.
- A dispenser handle can be set on the dispenser stand, on the main system, or even hanging onto other places to free up maximum bench space.
4. Final Filter
- A wide range of final filters (optional) is applicable to ensure ultrapure water without particles, bacteria or pyrogen.
5. Level sensor
- The tank integrated continuous level sensor measures water level within the tank and manages the system to start or stop producing water automatically based on requirements.
- Water level and quality inside the tank can be viewed from the control console.
Main Applications
Ultrapure Water- HPLC mobile phase preparation
- Preparation of reagent blank solution
- As sample diluents for GC, HPLC, UPLC, AA, ICP-MS and other analytical techniques
- Preparation of buffer and culture media for mammalian cell culture
- Preparation of molecular biology reagents, etc.
RO Pure Water
- Glassware cleaning
- Washing machine for glassware
- Water bath water
- Autoclave
- Feed water to laboratory animals
Specification
Feed Water Requirements | |
Tap water | Municipal water, Conductivity < 2000 μS/cm (1000 ppm) |
Operating temperature | 5 - 35 ℃ |
Feed water pressure | 1 - 6 bar (15 - 90 psi) |
Water Flow Rate | |
RO water production rate (@ 25℃) | 12, 24, 32 L/hr |
RO water dispensing rate | Up to 2.0 L/min |
Ultrapure water dispensing rate | Up to 2.0 L/min |
Product Water Quality | |
RO water conductivity (@ 25℃) | typically < 20 μS/cm (feed water conductivity < 500 μS/cm) |
RO rejection | 97 to 98% ionic rejection (new RO cartridge) > 99% organic rejection > 99% particulates & bacteria rejection |
Ultrapure water resistivity (@ 25℃) | 18.2 MΩ·cm |
Ultrapure water TOC* | < 2 ppb |
Particles in ultrapure water ( > 0.2 μm) | No particles (with a 0.2 μm final filter or RephiBio filter) |
Microorganisms in ultrapure water | < 0.01 cfu/ml (with a 0.2 μm final filter or RephiBio filter) |
Pyrogens (endotoxins) in ultrapure water | < 0.001 Eu/ml (with a RephiBio filter) |
RNase in ultrapure water | < 0.5 pg/ml (with a RephiBio filter) |
DNase in ultrapure water | <10 pg/ml (with a RephiBio filter) |
Product Water Quality Stored in Tank ( for tank recirculation models ) | |
Resistivity (@25℃) | > 1 MΩ·cm (default), adjustable |
Dimensions | |
Main system dimensions Width × depth × height | 32 × 44 × 54 cm (12.6 × 17.3 × 21.3 in) |
Dispenser dimensions Width × depth × height | 21 × 29 × 61 cm (8.3 × 11.4 × 24.0 in) |
Weight | 20 kg |
Input Voltage | 110 - 240 VAC |
Operating Voltage | 24 VDC |
Main system power | < 200 W |
*In the appropriate operating conditions, otherwise typically ≤5 ppb.
Ordering Information
- Genie U 12 Water System (RG0U01000)
- Genie U 12 Water System with TOC (RG0U010T0)
- Genie U 12 Water System, with tank recirculation (RG0U0100R)
- Genie U 12 Water System with TOC & tank recirculation (RG0U010TR)
- Genie U 24 Water System RG0U02000)
- Genie U 24 Water System with TOC (RG0U020T0)
- Genie U 24 Water System, with tank recirculation (RG0U0200R)
- Genie U 24 Water System with TOC & tank recirculation (RG0U020TR)
- Genie U 32 Water System (RG0U03000)
- Genie U 32 Water System with TOC (RG0U030T0)
- Genie U 32 Water System, with tank recirculation 9RG0U0300R)
- Genie U 32 Water System with TOC & tank recirculation (RG0U030TR)
View more about Genie U Water System on main site